Rice Lake 920i Installation Manual V3.10 User Manual

Page 30

Advertising
background image

24

920i

Installation Manual

SCALES Menu

Parameter

Choices

Description

Level 2 submenus

SCALEx

Allows configuration and calibration of each scale

CONFIG

Lists available and associated A/Ds

Level 3 submenus

GRADS

10000
1–9999999

Specifies the number of full scale graduations if SPLIT=OFF. (For multi-range and
multi-interval scales (SPLIT

OFF), the GRADS value is derived from the capacity and

display divisions specified for the range or interval.)

The value entered must be in the range 1–9999999 and should be consistent with legal
requirements and environmental limits on system resolution.

To calculate GRADS, use the formula: GRADS = Capacity / Display Divisions.

Display divisions are specified under the FORMAT submenu.

SPLIT

OFF
2RNG
3RNG
2INTVL
3INTVL

Specifies whether the scale is full-range (OFF), multi-range (2RNG, 3RNG), or multi-interval
(2INTVL, 3INTVL). For multi-range and multi-interval scales, see the submenu shown in
Figure 3-6 on page 31 and parameter descriptions in Table 3-4 on page 32.

FORMAT

PRIMAR
SECNDR
TERTIA
ROC

For standard scales (SPLIT=OFF), see Level 4 submenu descriptions in Table 3-3 on
page 28.

For multi-range and multi-interval scales, see Table 3-4 on page 32.

ZTRKBND

0
number

Automatically zeroes the scale when within the range specified, as long as the input is
within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display
divisions. Maximum legal value varies depending on local regulations.

NOTE: For scales using linear calibration, do not set the zero tracking band to a value
greater than that specified for the first linearization point.

ZRANGE

1.900000
number

Selects the range within which the scale can be zeroed. The 1.900000 default value is ±
1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use the default value for legal-for-trade applications.

MOTBAND

1
number

Sets the level, in display divisions, at which scale motion is detected. If motion is not
detected for 1 second or more, the standstill symbol lights. Some operations, including
print, tare, and zero, require the scale to be at standstill. Maximum legal value varies
depending on local regulations.

If this parameter is set to 0, the standstill annunciator is set continuously on, and operations
including zero, print, and tare will be performed regardless of scale motion. If 0 is selected,
ZTRKBND must also be set to 0.

SSTIME

10
number

Specifies the length of time the scale must be out of motion, in 0.1-second intervals, before
the scale is considered to be at standstill. Values greater than 10 are not recommended.

OVRLOAD

FS+2%
FS+1D
FS+9D
FS

Determines the point at which the display blanks and an out-of-range error message is
displayed. Maximum legal value varies depending on local regulations.

WMTTHRH

1000
number

Specifies the minimum number of grads required for a weighment to be added to the
recorded number of weighments.

Table 3-2. SCALES Menu Parameters

Advertising