4.0

Calibration

The

920i

can be calibrated using the front panel, serial commands, or

iRev

. Each method consists of the following

steps:

Zero calibration

Entering the test weight value

Span calibration

Optional five-point linearization

Optional rezero calibration for test weights using hooks or chains

The following sections describe the calibration procedure for each of the calibration methods. See the

iQUBE

Installation Manual, PN 77224, for information about configuring

iQUBE

-attached scales.

Figure 4-1. Calibration (CALIBR) Submenu

4.1

Gravity Compensation

Gravity compensation for latitude and elevation is available for the

920i

. To calibrate with gravity compensation,

the LOCALE parameter under the FEATURE menu must be set ON, and the LATUDE (latitude) and ELEVAT

(elevation, in meters) parameters set before calibrating the indicator (see Figure 3-11 on page 39).
If the indicator is later installed at a different location, gravity compensation can be applied to a pre-calibrated

indicator by adjusting the LATUDE and ELEVAT parameters.

4.2

Front Panel Calibration

The CALIBR submenu (under the SCALES menu,

see Figure 4-2) is used to calibrate the

920i

. The zero,

span, and linear calibration point displays provide a

set of softkeys used specifically for calibration

procedures:

+/–

Toggles to allow entry of negative or

positive values

Last Zero

Recalls the last established zero value

to allow calibration without removing

tests weights or product from scale.

Calibrate

Performs calibration for the selected

point

Temp Zero

Temporarily zeroes the displayed

weight of a non-empty scale. After span

calibration, the difference between the

temp zero and the previously calibrated

zero value is used as an offset.

Millivolts

(or

Counts

)

Toggles between display of captured A/

D counts and captured millivolts

values; allows entry of calibration

values in mV or counts

SCALES

SERIAL

FEATURE

PFORMT

SETPTS

DIG I/O

VERS

SCALE 1

CALIBR

WZERO

WVAL

WSPAN

WLIN

REZERO

POINT 4

POINT 5

POINT 1

POINT 2

POINT 3