Note

Note

Installation and Operation Manual - Supervisor Mode Parameters

41

There are many parameters in the

Supervisor Mode (shown in the following

sections), but most will not be used for a

basic installation. The parameters that

must be set prior to calibration include the following:
Calibration Weight (if using static weights to calibrate)

Idler spacing

Number of idlers

Belt length

Load cell MV

Total load cell build

The other parameters that are commonly set are the

filtering values but they can be done after setup and

calibration.

The angle of inclination is not required with

Rice Lake Weighing Systems frames due to

laterally opposed shearbeams.

4.2

Scale Capacity

This parameter defines the maximum rated capacity
for the belt scale or the maximum amount of weight
that can pass along the belt scale per hour.

The default parameter is measured in tons/hour.

Enter the maximum rate in tons using the numeric
keypad and press the

enter

key to save that parameter.

4.3

Load Cell MV

Enter the average mv/v rating of all the load cells in
the system.

Use the numeric keypad to enter that value which
must be greater than 0. Press the

enter

key to save that

parameter.

4.4

Total Load Cell Build

This parameter is the total load cell capacity of the all
the load cells in system. If you have eight load cells
for example and each one has a rated capacity of XX,
take that capacity and times it by 8 to get the total load
cell capacity to be entered into the integrator.

4.5

Rate Unit Time

This parameter defines what unit of time the rate will
be displayed in on the integrator. Available selections
are:

Hr. - hour (default)

Min - minute

Sec - seconds

Press the

enter

key to toggle between the listed

choices above. The default rate unit time is Hr.

4.6

Filter (s)

This parameter defines the number of seconds
filtering samples to average the scale load. It is
measured in tenths of seconds.

4.7

Filter Threshold (divisions)

This is the number os division the load must maintain
for the filter to be enabled. If the load sees a change in
weight that is greater than this value, the filter is
flushed.

4.8

Speed Unit Time

This parameter defines how the belt speed is displayed
on the integrator. The selections are displayed in the
following units.

Hr. - hour

Min - minute (default)

Sec - seconds

Press the

enter

key to toggle between the listed

choices above. The default speed unit time is
displayed in minutes.

4.9

Fixed Speed

This parameter allows a technician to configure a
fixed belt speed for the integrator. The integrator will
no longer calculate speed using the speed sensor.

The default value is 0 and if a value is set then the
integrator will used the fixed value. The unit of
measure is how the integrator has been configured for
speed so usually it will be either ft/min or m/min.

Note

Digital I/O six must be jumped to the

ground when using fixed speed in order

for fixed speed to work.

The user can change the time units to

be either seconds or hours. Enter a numeric value and
press the

enter

key to save that value.

4.10 Unit of Measure

The unit of measure parameter defines how the belt is
measured and speed calculated. The selections are
displayed in the following units.

Ft - feet (default)

M - meters

Press the

enter

key to toggle between the listed

choices above.