Zero Cal Calibration Mode

By selecting this mode will calibrate the integrator based on a zero test. A zero calibration is based on the number
of belt revolutions as established during the belt speed sensor calibration.

Use the following steps to perform a Zero Cal calibration.

1. Ensure that there is no material on the belt.

2. Press the

ZERO

key on the integrator. The previous zero information will be displayed with the zero scale

menu.

3. The operator will press the

Start

key.

4. The Zero Cal test will run, and after the test is completed the new zero error% will be shown on the

display. The operator needs to press either the

Yes

softkey to accept that value or

No

softkey to not see the

new zero.

10/31/2008

08:13AM

No

Test Time

Test Accumulator

Pulses

Yes

Zero % Error

13.50

117.4

3574

418.16

Accept New Zero

Figure 5-4. Zero Percentage Error Display

Auto Cal Mode

By selecting the Auto Cal mode, the integrator will calibrate the span using the number of calibration revolutions
as the reference for the calibration duration. Span calibrations are based on belt length defined by the number of
revolutions and use either static weights or test chains. Use the following steps to perform an auto calibration.

1. Press the

Auto Cal

softkey on the integrator display. The integrator will display

Test Weights

,

Test Chain

,

and

Exit

softkeys.

2. Select either the

Test Weights

softkey or the

Test Chain

softkey depending on the span calibration method

you are using.

Exit =>

Test

Weights

g

Test Chain

Figure 5-5. Zero Percentage Error Display

3. Load the scale with weights or chains, then press the

Start

softkey to initiate the calibration sequence.