3 serial command calibration – Rice Lake CB-2 Concrete Batch Controller Version 1.0 User Manual

Page 40

Advertising
background image

36

CB-2 Automated Concrete Batching System Installation/Operation Manual

To calibrate the indicator using the front panel, do the

following:

1. Place the indicator in setup mode (display

reads

Scale Configuration

) and remove all

weight from the scale platform. If your test

weights require hooks or chains, place the

hooks or chains on the scale for zero

calibration.

2. With the

SCALES

menu highlighted, press the

d o w n

key, then select the scale to be

calibrated. Press

d o w n

again (

G R A D S

parameter highlighted), then press

left

to

h i g h l i g h t t h e

C A L I B R

s u b m e n u ( s e e

Figure 5-1). Press

d o w n

to go to zero

calibration (

WZE RO

). See Figure 5-2 on

page 35.

3. Ensure scale is empty, then press

down

again

to show the current WZERO value, then press

the

Calibrate

softkey to calibrate zero. When

complete, the new A/D count for the zero

calibration is displayed. Press

enter

to save the

zero calibration value and go to the next

prompt (

WVAL

).

4. With

WVAL

displayed, press

down

to show the

stored calibration weight value. Use the

numeric keypad to enter the actual value of

the calibration test weights, then press

enter

to

save the value and go to span calibration

(

WSPAN

).

5. Place test weights on scale. Press

down

again

to show the current WSPAN value, then press

the

Calibrate

softkey to calibrate span. When

complete, the new A/D count for the span

calibration is displayed. Press

enter

again to

save the span calibration value and go to the

next prompt (

WLIN

).

6. Five-point linearization (using the WLIN

parameter) provides increased scale accuracy

by calibrating the indicator at up to five

additional points between the zero and span

calibrations.
Linearization is optional: if you choose not to

perform linearization, skip the WLIN

parameter; if linearization values have

previously been entered, these values are reset

to zero during calibration. To perform

linearization, follow the procedure below:
With

WLIN

displayed, Press

down

to go to the

first linearization point (

POINT 1

). Press

down

again to show the weight value prompt (

WGT

1

), then

down

once more to show the weight

value. Place test weights on the scale, then use

the numeric keypad to enter the actual test

weight value. Press

enter

to save the value and

move to the calibration (

CAL 1

) prompt. Press

down

to show the current calibration value,

then press the

Calibrate

softkey to calibrate the

linearization point. When complete, the A/D

count for the linear calibration is displayed.

Press

enter

again to save the calibration value

and go to the next prompt (

POINT 2

).

Repeat for up to five linearization points. To

exit the linearization parameters, press the

up

key to return to WLIN.

7. The optional rezero function is used to

remove a calibration offset when hooks or

chains are used to hang the test weights.

NOTE: The rezero function cannot be used with
five-point linear calibration.

If no other apparatus was used to hang the

test weights during calibration, remove

the test weights and press

up

to return to

the CALIBR submenu.

If hooks or chains were used during

calibration, remove these and the test

weights from the scale. With all weight

removed, go to the REZERO parameter,

then press

down

to show the current zero

value. Press the

Calibrate

softkey to adjust

the zero and span calibration values. Press

enter

or

up

to return to the CALIBR

submenu.

8. Press

up

to return to the SCALES menu, or

press the

Save and Exit

softkey to exit setup

mode.

5.3

Serial Command Calibration

To calibrate the indicator using serial commands, the

indicator serial port must be connected to a terminal or

personal computer. See Section 2.3.1 on page 9 for

serial port pin assignments.
Once the indicator is connected to the sending device,

do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Send the SC.WZERO#n serial command

(where n is the number of the scale) to

calibrate zero.

3. Place test weights on the scale and use the

SC.WVAL command to enter the test weight

value in the following format:

SC.WVAL#n=vvvvv<CR>

4. Send the SC.WSPAN#n serial command to

calibrate span.

5. Up to five linearization points can be calibrated

between the zero and span calibration values.

Use the following commands to set and