4 irev calibration – Rice Lake CB-2 Concrete Batch Controller Version 1.0 User Manual

Page 41

Advertising
background image

Calibration

37

calibrate a single linearization point:

SC.WLIN#n.V1=vvvvv<CR>
SC.WLIN#n.C1<CR>

The SC.WLIN#n.V1 command sets the test

weight value (vvvvv) for linearization point 1.

The SC.WLIN#n.C1 command calibrates the

point. Repeat using the SC.WLIN#n.Vx and

SC.WLIN#n.Cx commands as required for

additional linearization points.

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used

t o h a n g t e s t w e i g h t s , t h e n s e n d t h e

SC.REZERO#n serial command.

7. Send the KSAVE serial command to save the

calibration changes; send KEXIT to exit setup

mode.

5.4

iRev Calibration

The

iRev

Calibration Wizard provides step-by-step

scale calibration. With the

920i

connected to the PC,

select the Calibration Wizard from the Tools menu on

the

iRev

Scales display, then follow the steps listed

below to calibrate the scale. See Section 5.0 on

page 51 for a general overview of the

iRev

utility.

1. On the first Calibration Wizard display (see

F i g u r e 5 - 3 ) , s e l e c t w h e t h e r y o u a r e

performing a standard (zero and span)

calibration or a multi-point linear calibration.

Click the

Next

button to continue.

Figure 5-3. iRev Calibration Wizard

2. Next (see Figure 5-4), select the scale you

wish to calibrate.

Figure 5-4. iRev Scale Selection Display

3. Enter the test weight used to calibrate the

scale (see Figure 5-5). If chains or hooks are

used to hold the weights, check the box below

the test weight value entry. This adds a rezero

step to the calibration sequence.

Figure 5-5. iRev Test Weight Value Display

4. Remove all weights from the scale. If chains

or hooks are used, place them on the scale.

Press the

Calibrate Zero

button to perform the

zero calibration (see Figure 5-6). A message

box appears when the process is complete.

Figure 5-6. iRev Zero Calibration Display