1 configuration menu, Configuration menu – Rice Lake IQ Plus 355 Installation Manual V1.17 User Manual

Page 19

Advertising
background image

Configuration

15

3.2.1

Configuration Menu

Figure 3-5. Configuration Menu

CONFIG Menu

Parameter

Choices

Description

Level 2 submenus

GRADS

10000
number

Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–100 000 and should be consistent with legal requirements and environmental
limits on system resolution.

To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.

Display divisions for primary and secondary units are specified on the FORMAT menu.

ZTRKBN

OFF
0.5D
1D
3D

Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the configured zero range (ZRANGE parameter). Selections are ±
display divisions. Maximum legal value varies depending on local regulations.

Table 3-2. Configuration Menu Parameters

9999999

9999999

9999999

9999999