7.7

Digital Filtering

The IQ plus 355 uses averaged digital filtering to reduce the effect of vibration on weight readings. Adjustable

threshold and sensitivity functions allow quick settling by suspending filter averaging, allowing the weight

reading to jump to the new value. Figure 7-3 shows the digital filter parameters on the CONFIG menu.

Figure 7-3. Digital Filtering Parameters on the Configuration (CONFIG) Menu

Figure 7-4. Flow Diagram for IQ plus 355 Digital Filters

7.7.1

DIGFLx Parameters

The first three digital filtering parameters, DIGFL1,

DIGFL2, and DIGFL3, are configurable filter stages

that control the effect of a single A/D reading on the

displayed weight. The value assigned to each

parameter sets the number of readings received from

the preceding filter stage before averaging (see

Figure 7-4).
The overall filtering effect can be expressed by adding

the values assigned to the two filter stages:

DIGFLT1 + DIGFLT2

For example, if the filters are configured as

DIGFLT1=16, DIGFLT2=8, the overall filtering

effect is 24 (16 + 8 ). With this configuration, each A/

D reading has a 1-in-24 effect on the displayed weight

value. Setting the filters to 1 effectively disables

digital filtering.

7.7.2

DFSENS and DFTHRH Parameters

The three digital filters can be used by themselves to

eliminate vibration effects, but heavy filtering also

increases settling time. The DFSENS (digital filter

sensitivity) and DFTHRH (digital filter threshold)

parameters can be used to temporarily override filter

averaging and improve settling time:

DFSENS specifies the number of consecutive

scale readings that must fall outside the filter

threshold (DFTHRH) before digital filtering

is suspended.

DFTHRH sets a threshold value, in display

divisions. When a specified number of

consecutive scale readings (DFSENS) fall

outside of this threshold, digital filtering is

suspended. Set DFTHRH to NONE to turn off

the filter override.









%*('-















%*('-















%*('-