About this manual, 0 introduction, 1 operating modes – Rice Lake IQ Plus 355 Installation Manual V1.17 User Manual

Page 5: About this manual 1.0, Introduction, 7arning

Advertising
background image

Introduction

1

About This Manual

This manual is intended for use by service technicians
responsible for installing and servicing IQ plus

®

355

digital weight indicators. This manual applies to

indicators using Version 1.1 of the IQ plus 355

software. See Section 7.10 on page 45 for a summary

of software changes included in this release.
Configuration and calibration of the indicator can be

accomplished using the indicator front panel keys, the

EDP command set, or Version 2.5 or later of the
Revolution

configuration utility. See Section 3.1 on

page 12 for information about configuration methods.

Some procedures described in this
manual require work inside the
indicator enclosure. These procedures
are to be performed by qualified
service personnel only.

Authorized distributors and their employees

can view or download this manual from the

Rice Lake Weighing Systems distributor

site at

www.ricelake.com

.

The Operator Card included with this manual

provides basic operating instructions for users of the

IQ plus 355. Please leave the Operator Card with the

indicator when installation and configuration are

complete.

1.0

Introduction

The IQ plus 355 is a single-channel digital weight

indicator housed in a NEMA 4X/IP66-rated stainless

steel enclosure. The indicator front panel consists of a

large (.8 in, 20 mm), six-digit, seven-segment LED

display and five-button keypad. Features include:

Drives up to eight 350

or sixteen 700 load

cells

Supports 4- and 6-wire load cell connections

Two configurable digital inputs

Electronic data processing (EDP) port for full

duplex, RS-232 communications at up to

9600 bps

Printer port for output-only RS-232 and 20

mA current loop communications at up to

9600 bps

Optional analog output module provides 0–10

VDC or 4–20 mA tracking of gross or net

weight values

Available in 115 VAC and 230 VAC versions

The IQ plus 355 is NTEP-certified and Measurement

Canada approved for Classes III, III HD, and III L at

10,000 divisions. See Section 7.11 on page 46 for

detailed specifications.

1.1

Operating Modes

The IQ plus 355 has three modes of operation:

Normal (weighing) mode

Normal mode is the “production” mode of the

indicator. The indicator displays gross or net

weights as required, using the LED annunciators

described in Section 1.3 on page 2 to indicate

scale status and the type of weight value

displayed. Once configuration is complete and a

legal seal is affixed to the back of the indicator,

this is the only mode in which the IQ plus 355 can

operate.

Setup mode

Most of the procedures described in this manual

require the indicator to be in setup mode,

including configuration and calibration.
To enter setup mode, remove the large fillister

head screw from the enclosure backplate. Insert a

screwdriver or a similar tool into the access hole

and press the setup switch once. The indicator

display changes to show the word

CONFIG

.

Test mode

Test mode provides a number of diagnostic

functions for the IQ plus 355 indicator. Like setup

mode, test mode is entered using the setup switch.

See Section 7.9 on page 44 for more information

about entering and using test mode.

7ARNING