Rice Lake IQ Plus 355 Installation Manual V1.17 User Manual

Page 20

Advertising
background image

16

IQ plus 355 Installation Manual

ZRANGE

1.9%
100%

Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is
± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use 1.9% for legal-for-trade applications.

MOTBAN

1D
2D
3D
5D
10D
20D
OFF

Motion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected for 1 second or more, the standstill symbol lights. Some
operations, including print, tare, and zero, require the scale to be at standstill. Maximum
legal value varies depending on local regulations.

If OFF is selected, ZTRKBN should also be set to OFF.

OVRLOA

FS+2%
FS+1D
FS+9D
FS

Overload. Determines the point at which the display blanks and an out-of-range error
message is displayed. Maximum legal value varies depending on local regulations.

SMPRAT

15HZ
30HZ
3.75HZ
7.5HZ

Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital
converter. Lower sample rate values provide greater signal noise immunity.

DIGFL1
DIGFL2
DIGFL3

2
4
8
16
32
64
1

Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanical
vibration from the immediate area of the scale.

Choices indicate the number of A/D conversions that are averaged to obtain the
displayed reading. A higher number gives a more accurate display by minimizing the
effect of a few noisy readings, but slows down the settling rate of the indicator. See
Section 7.7 on page 42 for more information on digital filtering.

DFSENS

8OUT
16OUT
32OUT
64OUT
128OUT
2OUT
4OUT

Digital filter cutout sensitivity. Specifies the number of consecutive readings that must fall
outside the filter threshold (DFTHRH parameter) before digital filtering is suspended. If
NONE is selected, the filter is always enabled.

DFTHRH

NONE
2DD
5DD
10DD
20DD
50DD
100DD
200DD
250DD

Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When a
specified number of consecutive scale readings (DFSENS parameter) fall outside of this
threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.

TAREFN

BOTH
NOTARE
PBTARE
KEYED

Tare function. Enables or disables push-button and keyed tares. Possible values are:

BOTH:

Both push-button and keyed tares are enabled

NOTARE: No tare allowed (gross mode only)
PBTARE: Push-button tares enabled
KEYED:

Keyed tare enabled

CONFIG Menu

Parameter

Choices

Description

Table 3-2. Configuration Menu Parameters (Continued)