Configuration

21

3.2.6

Print Format Menu

See Section 5.0 on page 27 for information about custom print formatting.

Figure 3-10. Print Format Menu

PROGRM Menu

Parameter

Choices

Description

Level 2 submenus

PWRUPM

GO
DELAY

Power up mode. In GO mode, the indicator goes into operation immediately after a brief power
up display test.

In DELAY mode, the indicator performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset and
the warm up period repeated.

REGULA

NTEP
OIML
CANADA
NONE

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.

• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than

zero. NONE allows tares to be acquired at any weight value.

• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at

no load. NONE allows tares to be cleared at any weight value.

• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present.

In CANADA mode, the previous tare must be cleared before a new tare can be acquired.

• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net

mode as long as the current weight is within the specified ZRANGE. In OIML mode, the
scale must be in gross mode before it can be zeroed; pressing the ZERO key in net mode
clears the tare.

CONSNU

000000
number

Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.

The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.

CONSTU

000000
number

Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the indicator is powered on.

Table 3-6. Program Menu Parameters