EDP Command Calibration

To calibrate the indicator using EDP commands, the

indicator EDP port must be connected to a terminal or

personal computer. See Section 2.3.3 on page 6 for

EDP port pin assignments; see Section 6.0 on page 29

for more information about using EDP commands.
Once the indicator is connected to the sending device,

do the following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Send the WZERO EDP command to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight

value in the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress.

5. To remove an offset value, clear all weight

from the scale, including hooks or chains used

to hang test weights, then send the REZERO

EDP command. The indicator displays

*CAL*

while the zero and span calibrations are

adjusted.

6. Send the KUPARROW EDP command to exit

setup mode.

4.3

Revolution

Calibration

To calibrate the indicator using Revolution, the

indicator EDP port must be connected to a PC running

the Revolution configuration utility.
Use the following procedure to calibrate the indicator:

1. Select

Calibrate Indicator

from the Revolution

main menu.

2. On the Indicator Calibration display, select

t h e i n d i c a t o r m o d e l (

I Q + 3 5 5

) a n d

communications port then click

OK

.

3. Revolution uploads calibration data from the

indicator then presents the information in a

display like that shown in Figure 4-3.

Figure 4-3. Revolution Calibration Display

4. Enter the

Value of Test Weight

to be used for

span calibration then click

OK

.

5. The Zero Calibration dialog box prompts you

to remove all weight from the scale. Clear the

scale and click

OK

to begin zero calibration.

NOTE:

If your test weights require hooks or

chains, place the hooks or chains on the scale

for zero calibration.

6. When zero calibration is complete, the Span

Calibration dialog box prompts you to place

test weights on the scale for span calibration.

Place tests weights on the scale then click

OK

.

7. When calibration is complete, the

New Settings

fields of the Indicator Calibration display are

filled in. Click

Exit

to save the new values and

return to the Revolution main menu; to restore

the previous calibration values, click

Restore

Settings

.