IQ plus 355 Installation Manual

6.2

Saving and Transferring Data

Connecting a personal computer to the IQ plus 355

EDP port allows you to save indicator configuration

data to the PC or to download configuration data from

the PC to an indicator. The following sections

describe the procedures for these save and transfer

operations.

6.2.1

Saving Indicator Data to a Personal Computer

Configuration data can be saved to a personal

computer connected to the EDP port. The PC must be

running a communications program such as
PROCOMMPLUS

®

. See Section 2.3.3 on page 6 for

information about serial communications wiring and

EDP port pin assignments.
When configuring the indicator, ensure that the values

set for the BAUD and BITS parameters on the

SERIAL menu match the baud rate, bits, and parity

settings configured for the serial port on the PC. Set

the PRNDES parameter to EDP.
To save all configuration data, place the indicator in

setup mode and send the DUMPALL EDP command

to the indicator. The IQ plus 355 responds by sending

a l l c o n f i g u r a t i o n p a r a m e t e r s t o t h e P C a s

ASCII-formatted text.

6.2.2

Downloading Configuration Data from PC to
Indicator

Configuration data saved on a PC or floppy disk can

be downloaded from the PC to an indicator. This

procedure is useful when a number of indicators with

similar configurations are set up or when an indicator

is replaced.
To download configuration data, connect the PC to the

EDP port as described in Section 6.2.1. Place the

i n d i c a t o r i n s e t u p m o d e a n d u s e t h e P C

communications software to send the saved

configuration data to the indicator. When transfer is

complete, calibrate the indicator as described in

Section 4.0 on page 24.

NOTES:

Calibration settings are included in the

configuration data downloaded to the

indicator. If the receiving indicator is a direct

replacement for another IQ plus 355 and the

attached scale is not changed, recalibration is

not required.

When downloading configurations that

include changed serial communications

settings, edit the data file to place the serial

communications changes at the end of the

file. Communication between the PC and

indicator will be lost once the indicator

receives settings for baud rate (BAUD

parameter) or data bits and parity (BITS

parameter) that do not match those configured

for the PC.