IQ plus 355 Installation Manual

4.0

Calibration

The IQ plus 355 can be calibrated using the front panel, EDP commands, or the Revolution

configuration

utility. Each method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the

following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from the

scale platform. If your test weights require

hooks or chains, place the hooks or chains on

the scale for zero calibration.

2. Press until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration

(

WZERO

).

3. With

WZERO

displayed, press to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the zero calibration is

displayed. DO NOT adjust this value after
WSPAN has been set!
Press again to save

the zero calibration value and go to the next

prompt (

WVAL

).

4. With

WVAL

displayed, place test weights on

the scale and press to show the test weight

value. Use the procedure shown in Figure 4-2

to enter the actual test weight, then press to

save the value and go to span calibration

(

WSPAN

).

5. With

WSPAN

displayed, press to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress. When complete, the

A/D count for the span calibration is

displayed. Press again to save the span

calibration value and go to the next prompt

(

REZERO

).

6. The rezero function is used to remove a

calibration offset when hooks or chains are

used to hang the test weights.

If no other apparatus was used to hang the

test weights during calibration, remove

the test weights and press to return to

the CALIBR menu.

If hooks or chains were used during

calibration, remove these and the test

weights from the scale. With all weight

removed, press to rezero the scale. This

function adjusts the zero and span

calibration values. The indicator displays

*CAL*

while the zero and span calibrations

are adjusted. When complete, the adjusted

A/D count for the zero calibration is

displayed. Press to save the value, then

press to return to the CALIBR menu.

7. Press until the display reads

CONFIG

, then

press to exit setup mode.

Figure 4-2. Editing Procedure for Numeric Values

9999999

9999999

9999999

9999999

8;&30

*CAL*

%JTQMBZBOEFEJU

[FSPDBMJCSBUJPO

"%DPVOUWBMVF

%JTQMBZBOEFEJU

UFTUXFJHIUWBMVF

87"-

*CAL*

%JTQMBZBOEFEJU

TQBODBMJCSBUJPO

"%DPVOUWBMVF

841"/

7&34

"-(065

%*(*/

130(3.

1'03.5

4&3*"-

$"-*#3

$0/'*(

'03."5

CAL*

1SFTT&OUFSUP

SFNPWFPGGTFUGSPN

[FSPBOETQBO

DBMJCSBUJPOT

3&;&30

8IFOFEJUJOHOVNFSJDWBMVFT QSFTT

PS

UP DIBOHF UI F

EJHJUTFMFDUFE1SFTT

PS

UPJODSFNFOUPSEFDSFNFOU UIF

WBMVFPGUIFTFMFDUFEEJHJU

1SFTT

UPTBWFUIFWBMVFFOUFSFEBOESFUVSOUPUIFMFWFMBCPWF



Я



Я

ЯЯЯ 

Advertising