Brookfield DV2T Viscometer User Manual

Page 29

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 29

Manual No. M13-167-B0614

then rerunning the test (see Section III.8). This method can allow you

to quickly evaluate the spindle selection to determine the best speeds

for testing.

# of Points

: The test will complete when the specified number of data points has been

collected. Data is collected according to the Data Collection setting (see

Section III.5). The range of data points is: 1 – 9,999.

# of Revolutions

: The test will complete when the specified number of revolutions of the

spindle has occurred. Data is collected according to the Data Collection

setting (see Section III.5). The range for number of revolutions is:

1 – 9,999.

Torque

:

The test will complete when the specified Torque value is measured. Data

is collected according to the Data Collection setting (see Section III.5). The

range of values for the measured Torque is: -10.0 – 100.0.

Viscosity

:

The test will complete when the specified Viscosity value is measured. Data

is collected according to the Data Collection setting (see Section III.5). The

range of measured Viscosity is 0 – 10,000,000,000 cP.

Note: The Viscosity End Condition should be selected in consideration of

the Range provided by the spindle and speed selected. Check the Range

value by selecting Spindle in Configure Viscosity Test. If you choose a

Viscosity End Condition greater than the Range of the spindle and speed

settings, then the value shown in the End Condition field will be EEEE. In

this case, the end condition will never be achieved.

Temperature

:

The test will complete when the specified Temperature value is measured

to within the indicated Tolerance. The Tolerance specifies how close the

measured temperature should be to the specified temperature value to

consider the target reached. Temperature is measured through the use

of a connected Brookfield temperature probe (DVP-94Y, CPA-44ZPY

cup, SC4-XXRPY chamber). Data is collected according to the Data

Collection setting (see Section III.5).

Temperature: -100 – 300°C

Tolerance: 0.0 – 9.9°C

TIP: A small tolerance value will require a much longer time to reach

the End Condition.

TIP: If using PG Flash to create a test program, consider using a

larger Tolerance value for the End Condition and then use a

second step for thermal equilibrium (End Condition Time). This

two step method can reduce the variability in test time caused

by variations in temperature sensors.

III.7 Additional Test Parameters

The Configure Viscosity Test screen includes a More/Less Bar. When pressed, the More

Bar will present several additional Test Parameters including: QC Limits, Instructions,

Reports, and Density. These parameters are always available and active regardless of

the position of the More / Less Bar.