A Special Spindle is created by selecting the Special Spindle command button

in the Edit Spindle List menu. A Special Spindle requires a code, a name,

and an SMC value. Optional values include SRC and YMC.

Code

The code is the numerical value used to select the spindle. This

value must be unique and can not be the same as a standard

Brookfield spindle. The acceptable range for Code is 100-199.

Name

The Name is a unique value to describe the spindle. The Name

will be displayed in the Spindle field. The Name can be comprised

of alpha and numeric characters.

SMC

The SMC (Spindle Multiplier Constant) is used to convert the

measured torque to viscosity. The acceptable range of SMC is

0.001 – 500.
The SMC value can be determined through the use of the Range

equation found in appendix D. Some experimentation with the

spindle and a calibrated viscosity standard is required.
Range (cP) = TK * SMC * 10,000/RPM
The Torque Constant (TK) can be found in Table D-2 in Appendix D.

SRC

The SRC (Shear Rate Constant) is used to convert the set RPM

to shear rate and calculate Shear Stress. The acceptable range

of SRC is 0.001 – 500.
The SRC value can be determined if the shear rate characteristics of

the spindle geometry are known. Consult the Brookfield publication,

“More Solutions to Sticky Problems”, for a discussion about shear

rate and some recommended equations.
Shear Rate (1/s) = SRC * RPM

YMC

The YMC (Yield Multiplier Constant) is used to convert the measured

torque to yield stress.

This feature is only available with the DV3T Rheometer. Consult
the DV3T operating instructions for more information.

Speed List:

The selection of Speed in Configure Viscosity Test can be done by using either

a number pad or a scroll list. (This choice is selected through the navigation

button in the Set Spindle screen.) When scroll list is selected, the speed

entry is restricted to the classic Brookfield speed sets for the LV and RV/HA/

HB viscometers. The contents of the scroll list are defined in the Speed List

menu. Each speed can be individually selected or deselected through the use

of a check box. Select or deselect all speeds by using the Select All check

box. Select or deselect all LV (or RV/HA/HB) speeds by using the Select All

LV check box. The classic Brookfield speed sets are defined below.

LV:

0.3, 0.6, 1.5, 3, 6, 12, 30, 60

RV/HA/HB:

0.5, 1.0, 2.0, 2.5, 4, 5, 10, 20, 50, 100

Additionally, a new speed can be added to the list by selecting the New Speed

command key at the bottom of the display. The New Speed must be selected

within the range of available speeds on the DV2T: 0.0 – 200 RPM.
Speed entry through the number pad is never restricted.

TIP: To ensure that the correct speed is always entered, create a test that

is saved to the DV2T memory.