Brookfield DV2T Viscometer User Manual

Page 47

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 47

Manual No. M13-167-B0614

TIP: The Global Alarm sound can be turned off in the Sounds menu

of User Settings.

The Global Alarm is deactivated by choosing None in the drop down list of

available measurement parameters. You must press OK to confirm your

Global Alarm choice.

Spindle List:

The selection of Spindle in Configure Viscosity Test can be done by using

either a number pad or a scroll list. (This choice is selected through the

navigation button in the Set Spindle screen.) The contents of the scroll

list are defined in the Spindle List menu. Each spindle can be individually

selected or deselected through the use of a check box. Spindles, shown

with both their name and their number, are listed numerically according

to their spindle entry number.

The restricted spindle will not appear on the scroll list of available spindles

when choosing Spindle in Configure Viscosity Test. If the user selects to

enter the spindle by number, then the choice of a restricted spindle will

present an Invalid Input message and prevent the selection of the spindle

(see Figure IV-8).

List Button

Figure IV-8

TIP: To ensure that the correct spindle code is always entered, create a

test that is saved to the DV2T memory.

TIP: The selection of spindle can be restricted by User Log In. Use this

restriction combined with a saved test configuration to ensure that

the correct spindle is always used for critical tests. See Section

IV.4.2 for details of Users and Access.

A Special Spindle can be created when the user has developed a new spindle

geometry or when using a standard Brookfield spindle in a non standard

measurement container. The Special Spindle, once created, will be available

to users through the Spindle selection in Configure Viscosity Test.