Date/Sample filters

Complete selection of Date/Sample filter options per the instructions
in Chapter 6.

NOTE

The VI Matcher does not match invalid samples UNLESS the Allow
Matching ... box from the Sample Filter tab options is checked. The VI
Matcher never matches unsuccessful samples (samples with no flow
time and a displayed kinematic viscosity of “0”). If one of the samples ran
successfully at one temperature but not at the other, the good sample
can still be used for VI matching if the failed sample is rerun within 8
hours.

VI Analysis filter

To adjust other VI analysis options, click the

VI Report

tab.

VI Report data filter options

Then use the check boxes to select which fields to include in the
report. After individual options have been selected, use the related

spin controls

and radio buttons

to determine the appearance of

data for kinematic viscosity, temperature and drop times.

8. When you have completed the configuration, click

OK

. The program

will prompt you to save the configuration.