Checking wash and advanced settings – Cannon Instrument miniAV-X User Manual

Page 20

Advertising
background image

CANNON

®

miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.Ob— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

14

The Calibration window

Compare this value to your archive of the calibration constant data (if
available) for bulb 1. The values should be identical.

3. If they are not, place your cursor in the appropriate field, delete the

entry, then type the correct values for the constant in the text box.

4. Click

Update tube

.

5. Use the bulb spin controls

to select the other bulb(s) for which

you noted calibration constant errors. Input the correct values for
each. Make certain to click

Update Tube

after you have corrected

C

and

E

calibration values for each bulb BEFORE selecting the next

bulb.

6. When you have entered corrected constant values for each bulb on

both tubes, click

to exit the

Calibration

window.

You have verified the software configuration of VISCPRO

®

. To test

samples with your instrument, follow the instructions in Chapter 2. For
additional details regarding operating procedures for your instrument or
software, consult the appropriate section of this manual.

Checking Wash and Advanced Settings

To check Wash Settings for your miniAV-X instrument, click

Configure

from the VISCPRO menu options and select the desired instrument. Then
click

Tray Settings: Wash

. The

Tray Settings: Wash

window will

appear.