Cannon Instrument miniAV-X User Manual

Page 65

Advertising
background image

59

CANNON

®

miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0b— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Date selection:

The calibration program uses data from previously-run

standards to determine the calibration constants. To select the desired

standard data, use the

No older than [x] days

spin controls

to set the

date parameters for the database search (if “

1

” is selected, only the

standards run in the past 24 hours will be used).

Bulb selection:

Viscometer bulbs are numbered from the bottom to the

top of the viscometer tube. Use the spin controls to select the correct bulb
for calibration.

Data selection:

Test data from standards run with the miniAV-X within

the specified

Date

parameter will be automatically displayed in the list

box.

To select and deselect the data from the sample runs to be used for the
calibration, press and hold the

C key and click on the samples to

highlight those you wish to include in calibration calculations.

NOTE

Sample run data from at least two different standards for each bulb at
each calibration temperature must be available for calibration calcula-
tions to be performed.

After you have selected the desired data, release the

C key and click

Calculate new constants

to update values for C and E.

You may evaluate the accuracy of the new calibration constants by
comparing the known viscosity for the standard with the calculated
viscosity for each standard selected for the new calibration. This differ-
ence is displayed as a percentage in the last column (% Diff) of the
updated sample list box.

Show Calibration Data

The

Show Calibration Data

button is a toggle that will permit you to

compare current and new calibration data and constant information. Press
the

Show Calibration Data

button once to view the current calibration

information. Note that the button name will change to

Display Sample

Data

. Press the button again to display the new calibration data.

When/if you are satisfied with calculation results, click on

Update Tube

to save calibration values for the current configuration. The new data will
become the current data.

If the difference is still too great, repeat the calibration process until the
difference falls within acceptable limits (refer to the precision statement
in ASTM D445).