miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0b— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

When the

Wash

cycle has been completed, click

Done

to exit the

Wash

viscometer tube

window.

NOTES

The Wash operation cannot be executed if the tube is currently running a
sample. To terminate the Wash cycle at any time, click
Abort Wash.

CAUTION

The Wash viscometer tube function overrides automatic miniAV-X
software operations. Verify that the tubes are clean and dry before
initiating computer-controlled sample testing.

To adjust the vacuum/pressure settings for the miniAV-X, click

Service

from the VISCPRO menu options and select the desired instrument group
and instrument. Then click

Adjust Vacuum/Pressure

to open the

Adjust

Vacuum/Pressure

window.

Select the desired adjustment (vacuum or pressure) and then follow the
screen prompts to complete the adjustment procedure, using the appropri-
ate controls on the instrument wing panel. The prompts will display
recommended vacuum/pressure settings and will also indicate how
adjustments may be made to alter the settings.