Measuring kinematic viscosity, Safety cautions – Cannon Instrument miniAV-X User Manual

Page 8

Advertising
background image

CANNON

®

miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.Ob— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

2

Measuring kinematic viscosity

Kinematic viscosity is a measure of the internal resistance to flow of a
fluid under gravity with the pressure head being proportional to the
density of the fluid. For any particular viscometer, the time of flow of a
fixed volume of fluid is directly proportional to its kinematic viscosity.

Units of measure

An accepted unit of kinematic viscosity is one centimeter squared per
second, which is called one stoke. The centistoke (which is equivalent to
1 mm

2

/s) is the unit of measure most frequently used.

Methodology

ASTM Methods D445 and D 446, included with this manual, describe
appropriate test methodologies and instruments for glass capillary
viscometry.

Manual viscometers

Sections 9-11 of ASTM D445 provide detailed instructions for using
manual viscometers. ASTM D 446 suggests a minimum flow time of 200
seconds for nearly all the glass capillary viscometers (see tables in ASTM
D 446).

Automatic viscometers

For automatic viscometers, ASTM D445 Section 6.1.2 states, “Auto-
mated apparatus may be used as long as they mimic the physical condi-
tions, operations or processes of the manual apparatus they replace ... The
automated apparatus shall be capable of determining kinematic viscosity
of a certified viscosity reference standard within the limits stated ...”

Thus, automated viscometers can be used with flow times less than 200
seconds, as long as the kinetic energy correction and precision require-
ments are met.

miniAV-X tube characteristics Each standard miniAV-X viscometer tube has two bulbs, each of which

has its own calibration. The normal flow times for each bulb are 40-400
seconds. Each tube has a hundredfold measurement range (for example,
range from 2-200 cSt or 20-2000 cSt for each tube).

NOTE

CANNON

®

Instrument Company has not recommended the use of longer

flow times with the miniAV-X, as shorter flow times allow greater produc-
tivity. With longer flow times, the sample throughput would be signifi-
cantly reduced. However, the viscometer and software design does
permit longer efflux times (up to 600 seconds) as desired by the user.

Safety cautions

Please observe the following safety procedures and notices for proper
operation of the miniAV-X:

Do NOT lift the unit by the side panel “ears”. Always support the
unit from the base when moving or lifting.

Advertising