Ventilation, Checking bath temperature, Dual-solvent washing – Cannon Instrument miniAV-X User Manual

Page 87: Solvent wash by computer

Advertising
background image

81

CANNON

®

miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.0b— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Ventilation

The miniAV-X requires installation of the supplied 3” ventilation hose to
the exhaust shroud on the rear panel of the unit. The shroud must be
connected to a customer-supplied exhaust fan/ventilation system. A 3”-4”
hose adapter is shipped from CANNON Instrument Company with each
miniAV-X unit. The air flow rate and design of the exhaust fan must be
suitable to handle the vapors from the customer-supplied solvent and
samples.

Checking bath temperature

After the miniAV-X has been turned on, you may monitor bath tempera-
ture via a reference thermometer in the bath or via the software tempera-
ture display (Sample Input window). Temperature should rise until the
bath has attained the desired test temperature. Consult the reference
thermometer to ensure that the miniAV-X is controlling at the desired
operating temperature. If the computer temperature display does not
agree with the reference thermometer, perform the temperature calibra-
tion procedure per manual instructions.

Dual-solvent washing

For testing heavy oils or hard-to-clean materials,

CANNON

®

Instrument

Company recommends a dual-solvent washing system. The dual-solvent
washing process uses two different solvents. The first wash might
ordinarily use a heavy solvent to clean the sample residue out of the
viscometer tubes. The second wash might use a lighter solvent to clean
the remaining sample and solvent residue from the viscometer tube.
Consult Chapter 1 for information on dual-solvent connections.

Solvent wash by computer

Use the following procedure for computer-controlled solvent washing:

Procedure

1. If the desired tube is testing a sample, wait until the test is complete

OR click the

Abort

button from the

Sample Input wi

ndow and

select the tube/tray to abort from the

Select Trays

window. Then

click

OK

.

2. Proceed with the solvent wash by clicking the

Wash Tubes

button

from the

Sample Input

or

Machine Status

windows. Or you may

click on

Service/Wash Viscometer Tube

from the VISCPRO

®

primary menu options.

3. Click on

Begin Wash

to initiate a solvent wash using current wash

settings. During the wash cycle, the

Begin Wash

button will change

to

Abort Wash

. If it is necessary to interrupt the wash cycle, click

Abort Wash

.

NOTE

The wash procedure will have to be successfully completed before
further sample testing will be allowed.

Advertising