The entered actual temperature must not cause a correction in tem-
perature greater than +/- 5.0 degrees.

Click the Service/Zero Temperature Offsets menu item and select

which

offsets to clear.

Multiple corrections

Multiple or successive entries of correction for any temperature are
permissible. A newly entered corrected temperature is mathematically
compensated with the previously entered correction and is stored in
memory by the onboard microprocessor as the latest correction.

Data entry errors

The miniAV-X will not accept temperature reference data outside the
parameters noted previously. If the amount of the correction (delta to
displayed temperature) is too large, the correction will not be accepted.

General/Current Offsets

The

General Bath Offset

is factory-preset value affecting temperature

control at all temperatures and should not ordinarily be changed. The

Current Offset

is the temperature adjustment applied to a particular

temperature to conform bath temperature control to your temperature
reference . Always click the radio button for the

Apply As Current

Option

when calibrating for a specific temperature.

Training sensors

It is necessary to train instrument sensors in
preparation for sample testing at a specific
temperature. The training procedure estab-
lishes the proper power levels for sensor
detection of the sample meniscus. The sensor
training data is then stored by the miniAV-X
instrument, and recalled when testing at that
temperature. The sensor training procedure
requires approximately 30 minutes (longer at
lower temperatures).

NOTE

You will be prompted to train sensors by the
instrument hardware or software whenever a
new temperature is selected for miniAV-X
operation. You will be unable to test samples
until the training has been completed.
Sensor training data remains valid until the sensors are retrained at the
same temperature. Retraining is necessary only when the viscometer
tube is changed, or if sensor performance changes over time.

Train Tube Sensors

To train tube sensors, click

Service

menu and select the

Train Tube

Sensors

menu for the desired instrument. The

Train Tube Sensors

window will open.

Prewash option:

Click the

Prewash

option if the tube is not clean and

dry. VISCPRO will initiate a cleaning cycle using current wash configu-
ration settings prior to initiating the sensor training procedure.