Viewing/editing setup information – Cannon Instrument miniAV-X User Manual

Page 18

Advertising
background image

CANNON

®

miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.Ob— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

12

Viewing/editing setup information

NOTE

For first-time installation, make certain that the factory-prepared Configu-
ration file has been copied to the Default Data directory per instructions
on page 7 of manual. If your instrument has already been configured,
you can use the instructions in this section of the manual to check or, if
necessary, change the instrument settings.

1. Click

Configure

from the VISCPRO

®

menu bar.

2. Select your instrument group and instrument from the list of avail-

able instruments.

3. Select

Tray Settings: Tube and Bath

from the list of configuration

options.

The Tray Settings: Tube and Bath

window will appear.

The Tray Settings: Tube and Bath window

The

Tray Settings: Tube and Bath

window contains setup information

for each tube associated with your instrument. On multiple-tube CAN-
NON instruments, you can click on the tube tabs to see the setup infor-
mation for different tubes.

4. Verify that the tube serial number (

Tube S/N

) is correct. If it is not,

input the correct serial number in the text box. Each tube should have
a unique serial number!

5. Verify the presence of calibra-

tion values (

C

and

E

) for each

bulb. If calibration data is not
available, the default values
are C=1 and E=0.

Sample calibration constants