Soak time

Input the desired sample soak times for each bulb of the viscometer tube.
The soak time is the length of time in which the sample will remain in the
viscometer tube before being dropped. The soak time is important to
ensure temperature equilibration of the sample. The range is from 0 to
240 based on which units of time the user selected from Soak Time Units.

Constants

The

Tray Settings: Tube and Bath

screen displays the calibration

constants (

C

and

E

) for each bulb of the viscometer tube. These values

cannot be edited from the

Tray Settings . . .

screen. For information on

determining bulb constants for viscometer tubes, see Calibration later in
this chapter.

NOTE

When you have completed the configuration and saved the current tray
settings, you may wish to permanently store the current instrument
settings using the
Save Instrument and Tray Settings item in the Config-
ure
menu for a given instrument.

Tray Settings: Test

The

Tray Settings: Test

window is used to select parameters for

miniAV-X sample testing. Each sample drop is called a determination.
Multiple determinations are possible during testing of a single sample.
Comparisons of sample drop results permit user regulation of test preci-
sion.

The

Maximum Absolute Difference

option permits you to input the

maximum acceptable deviation for multi-drop samples, expressed as the
difference in the sample drop times in seconds.

The

Maximum % Difference

option allows you to input the maximum

acceptable deviation for multi-drop samples expressed as a percentage
(0.35 is the default percentage difference allowable). Values from 0.01 to
99.99 are acceptable. If the difference between test drop times is greater
than the acceptable deviation, the sample will be tested again until the
deviation in drop times is less than the

Maximum % Difference,

or until

the maximum number of determinations has been reached.

The

Maximum Determinations

option permits you to select from 1-20

iterations of the test using the same sample.

Tolerance parameters

Determine if the tolerance for differences between sample run times
(determinability) is going to be decided by the absolute difference in drop
time values (in seconds) or by the percentage difference between the drop
times. Click on the appropriate check box to select your choice(s).