Cannon Instrument miniAV-X User Manual

Page 68

Advertising
background image

CANNON

®

miniAV-X Automatic Viscometer with VISCPRO

®

Instruction & Operation Manual

Version 1.Ob— March, 2012;

CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

62

To calibrate for temperature, see Chapter 3 for details. Temperature
calibration requires a reference thermometer with an accuracy meeting
the precision specification of ASTM D445.

To zero temperature offsets, click the

Service/Clear Temperature

Offsets

menu option for the desired instrument. The

Zero Temperature

Offsets

window will open. Click the check box to select each desired

action. Then click

OK

to clear the selected settings.

If the

Clear Current Offset

option is checked, VISCPRO will clear the

calibration point for the current bath temperature.