The Calibration window

NOTE

You must click Update tube before changing bulbs if you wish to save
the new calibration constants for that bulb. If you close the
Calibration
window without updating the tube, the new calibration information will not
be saved.

Service menu

To access the service menu items, click

Service

from the VISCPRO

®

primary display and select the desired instrument.

The

Monitor

window permits the user to view current operational values

for the miniAV-X instrument. These settings may be helpful to

CAN-

NON

®

technical personnel when troubleshooting difficulties with the

miniAV-X.

Select

Wash viscometer tube

to open the

Wash viscometer tube

window.

Click

Begin Wash

to initiate the wash cycle. The

Status:

indicator will

display and update current actions.

Bypass .... option

To begin washing an evacuated sample cup, click the option box to make
certain that the

Bypass initial cup evacuation

option is selected.

CAUTION

Do not use this option with a filled cup or the cup will overflow.