5 Operation of NucleoCounter SP-100

29

time an analysis is performed. The Analysis Number can be reset to 0 using the
resetting function.

This feature is only used when a printer is connected to the instrument. Please refer to

Technical Note No. 100

with respect to using this feature.

To reset the Counter type “

F30+

”:

Display Text Type 26

Reset counter?

Reset counter?

Reset counter?

Reset counter?

Continue

Continue

Continue

Continue

Esc Cancel

Esc Cancel

Esc Cancel

Esc Cancel

Figure 54. Press “↵

” to reset the counter or press the “Esc” to cancel the operation.

If “↵

” is selected the display will shortly show "Resetting..." indicating that the Analysis

Number is being reset to 0:

Display Text Type 27

Resetting...

Resetting...

Resetting...

Resetting...


Figure 55. The counter is being reset.

5.15 Print to a PC (F310, only relevant when a PC is connected to printer port)

The function

F310 (Print -> PC) is used to choose between a normal print report to a

printer or a short print report designated for a PC. The default setting is “

Print -> PC

Off” and is used when a printer is connected to the instrument. The “Print -> PC On” is
designated for printing/logging the result of the analysis to a PC via the RS-232 printer
port. Please refer to

Technical Note No. 100

with respect using this feature.

Use the F310 command to to choose the type of print report:

Display Text Type 28

Print

Print

Print

Print-

-

-

->PC

>PC

>PC

>PC

Off

Off

Off

Off

1:On

1:On

1:On

1:On

2:Off

2:Off

2:Off

2:Off

store

store

store

store

Figure 56. Use the “1” and “2” buttoons to toogle between On and Off. Press “↵

” to

save the chosen setting or press the “Esc” to cancel the operation.