8 Troubleshooting, Error messages

41

8 Troubleshooting, Error messages

Under certain conditions the NucleoCounter SP-100 will display error messages during
operation. Corrective actions are suggested below but if they do not correct the errors,
contact ChemoMetec A/S or your local distributor.

With respect to problems with the transfer of images and results to the PC via the
USB, please refer to the User’s Guide for the SemenView software.

8.1

No valid cassette

When trying to analyze an SP1-Cassette, which is not valid, the NucleoCounter SP-100
displays the error message shown in

Figure 70

.

Display Text Type 33

No valid

No valid

No valid

No valid

cassette

cassette

cassette

cassette


Press an

Press an

Press an

Press any key

y key

y key

y key

Figure 70. The error message displayed when the NucleoCounter SP-100 does not

recognize the cassette as valid

This error message is displayed in the following situations:

The SP1-Cassette has been analyzed previously

The SP1-Cassette is not inserted properly into the NucleoCounter SP-100.

There is no SP1-Cassette inserted into the NucleoCounter SP-100.

Press any key to go to the Ready status.

8.2

Analysis aborted

The NucleoCounter SP-100 displays the error message shown in

Figure 71

when the

analysis has been aborted because the stained mixture inside the SP1-Cassette has not
reached the chamber within a given time.

Display Text Type 34

Analysis aborted

Analysis aborted

Analysis aborted

Analysis aborted

Flow error

Flow error

Flow error

Flow error

Refer to manual

Refer to manual

Refer to manual

Refer to manual

Press an

Press an

Press an

Press any key

y key

y key

y key

Figure 71. Error message displayed by the NucleoCounter SP-100 when the analysis has

been aborted.