6 Description of a typical Analysis Procedure (Total Count)

35

6 Description of a typical Analysis Procedure

(Total Count)

The Application Notes describe the analysis procedures in detail (e.g.

Application Note

No. 100

deals with the analysis of boar ejaculates).

Always refer to the latest version of the appropriate

Application Note

.

In this section a brief description of

Application Note No. 100 Version 1

is given. The

individual procedural steps are illustrated with pictures.

The following list contains the items required for determing sperm cell density in boar
semen with the NucleoCounter SP-100 system.

NucleoCounter SP-100 with software version 1.04 or higher

SP1-Cassettes

Container with Reagent S100 (lysis/dilution buffer)

20 ml polyethylene sample cups with polypropylene lids

Container holder and sample cup holder

Bottle-top dispenser 1-10 ml, e.g. 1-10 ml Brand Dispensette III Variabel

dispenser equipped with a 298 mm filling tube

Adjustable, automatic pipette, e.g. 10-100 µl Finnpipette adjusted to give 50 µl

boar semen ejaculate, and pipette tips

Optionally the NucleoCounter SP-100 can be connected to a computer via an

USB-interface. When connected to a computer the SemenView

TM

software offers

various features such as documentation of the obtained results.