6 Description of a typical Analysis Procedure (Total Count)

37

4.

Immediately, i.e. max. 10 seconds after the dispensing step, a portion of the
lyzate mixture shall be aspirated into a SP1-Cassette (see

Figure 65

). The tip of

the cassette should be immersed below the surface of the sample during
aspiration. Apply a consistent pressure to the piston and press the piston all the
way until it reaches the finger grips of the cassette.

Avoid touching the

window (clear area) of the measurement chamber....

Figure

Figure

Figure

Figure 65

65

65

65.

.

.

. Loading the cassette with lyzate mixture

5. Once the sample has been aspirated into the SP1-Cassette, open the lid and

insert the cassette in the insertion slit of the NucleoCounter SP-100. Close the lid
and press the “

Run” button on the instrument in order to initiate the analysis

(see

Figure 66

).

Figure

Figure

Figure

Figure 66

66

66

66.

.

.

. Insertion of cassette (A), closing the lid (B) and pressing the “

Run” button (C)

A

B

C