6 Description of a typical Analysis Procedure (Total Count)

36

Briefly, the Total Count analysis procedure comprises the following steps:

1. Type in the Sample ID on the NucleoCounter SP-100 and check that the Dilution

Factor is correct.

2. Using the automatic pipette mounted with a new pipette tip aspirate 50 µl of

the ejaculate. Transfer the sample to the center area of the bottom of a 20 ml
sample cup (see

Figure 63

).

Figure

Figure

Figure

Figure 63

63

63

63.... Pipetting of 50 µl of sample (A) and transferring sample to sample

cup (B)

3. Place the sample cup in the cup-holder and dispense 10.00 ml of Reagent S100

into the cup using the dispenser (see

Figure 64

). Dispense the reagent directly

onto the semen sample at the bottom of the cup. Use a consistent plunger
pressure, so that the plunger moves smoothly. There is no need for further
mixing, since the sample and the reagent are thoroughly mixed during the
dispensing of the Reagent S100.

Figure

Figure

Figure

Figure 64

64

64

64.... Placing the sample cup in the cup holder (A), dispensing 10 ml of

Reagent S100 (B) and sample cup with lyzate mixture (C)

B

A

C

B