ChemoMetec SP-100 User Manual

Page 64

Advertising
background image

9 Technical specifications

46

analysis is presented in

Millions of Cells per ml

of the initial Sample. The result is

always shown on the NucleoCounter display. If connected to a PC with
NucleoView installed the result is also displayed on the PC monitor. The viability
in percent can also be displayed on the PC-monitor if both a Total Count and a
Non-Viable Count are performed.

Operation

Menu-controlled by means of keyboard and LCD display.

Analysis time

When pressing “Run

Run

Run

Run” on the NucleoCounter SP-100 the result will be displayed

within:

30 seconds for a Total Count

60 seconds for a Non-Viable Count

Physical data

Weight

Height

Width

Depth

3 kg

26 cm

38 cm

22 cm

Power

See Section 11.5 Power Supply

Input

Fuse

12VDC (11-13VDC)

UL Listed fuse: 600mA/250Vac (UL 248)

Power consumption
NucleoCounter

Peak

Ready mode

Standby

25 W

2.5 W

2 mW

Operation conditions

Maximum relative humidity 80 percent for temperatures up to 31°C decreasing
linearly to 65 percent relative humidity at maximum 35°C; minimum temperature
15°C.

USB (For image and result
transfer to PC)

USB, version 1.1.

Note: Does not support USB Hubs

RS-232 Printer port

(To be used for printer,
SW-updating and result
transfer to PC)

Baud rate:
Data bits:
Parity:
Stop bits:
Flow control:

19200
8
None
1
None

Other environmental operating conditions for the NucleoCounter SP-100:

Indoor use only

Altitude

Up to 2000 m

Mains supply

Refer to the specifications for the Power supplies in section 11.5 Power supply

Installation Category

II (Refer to UL 61010A-1)

Overvoltage Category

II (Refer to UL 61010A-1)

Pollution Degree

II (Refer to UL 61010A-1)