Hardy HI 1769-FC User Manual

Page 49

Advertising
background image


44

Chapter 4

Mode

Value

RATE_CAL_MODE 7
REFILL_MODE 8
FULL_MODE

9

MANUAL_HOLD_MODE

10

RATECAL_HOLD_MODE 11

Integrated Technician

If there is NO IT JBOX, the instrument can still measure the sense voltage, the overall
load cell millivolts per volt, and the internal reference weight value. If the user has an IT
JBOX, the instrument can also read the weights and voltages of the individual sensors
attached to the IT JBOX.

To initiate an IT test, do a WRITEPARAM command, with parameter number 0x0036,
and a parameter value equal to the number of sensors, which should be 1-4 if you have an
IT JBOX.

To read back the results, do a READPARAM command, with parameter numbers 0x0036
through 0xB036. The return values are listed below.

Integrated Technician Return Values Table

Return Value

# Hex

Description

PARAMID_ITECH xx36

INSTANCE_VSENSE

0036 Sense voltage, 2 decimal places

INSTANCE_DVM_COMBINED

1036 millivolts per volt, 4 decimal places

INSTANCE_IREF_WEIGHT

2036

internal reference weight, weight
decimal point

INSTANCE_JBOX_MV/V1

3036

Load cell 1 millivolts per volt, 4
decimal places

INSTANCE_JBOX_MV/V2

4036

Load cell 2 millivolts per volt, 4
decimal places

INSTANCE_JBOX_MV/V3

5036

Load cell 3 millivolts per volt, 4
decimal places

INSTANCE_JBOX_MV/V4

6036

Load cell 4 millivolts per volt, 4
decimal places

INSTANCE_JBOX_REF_WEIGHT

7036

reference weight on IT JBOX,
weight decimal pt

INSTANCE_JBOX_WEIGHT1

8036

Sensor 1 on JBOX, weight decimal
point

INSTANCE_JBOX_WEIGHT2

9036

Sensor 2 on JBOX, weight decimal
point

INSTANCE_JBOX_WEIGHT3

A036

Sensor 3 on JBOX, weight decimal
point

INSTANCE_JBOX_WEIGHT4

B036

Sensor 4 on JBOX, weight decimal
point