G - Calibration Failed: A to D and Motion Errors

You may ZERO out chains and temporary calibration equipment to hold or hoist test
weights. Zeroing the temporary weight does not affect the calibration.

H - Mechanical Inspection

G

Verify some base operational conditions

Using a millivolt meter, verify that:

• The signal millivolt reading is positive and within the
acceptable range of 0 to 15 mV.

• When weight is applied there is a positive increase
in the signal millivolt readings

• The compression load cell has not been installed
upside down. Refer to the load cell installation guide
for installation instructions.

• No mechanical binding is restricting the vessel
movement during the load

• The load cell wiring is correct

• Each load cell signal changes under load. A single
l

oad cell installed upside down or wired backwards

can change the total signal output.

ERROR

PROCEED
WITH

CALIBRATION

Contact Hardy

Customer Service

(800) 321-5831

Ext. 1757

Yes

No