Chapter 7 – Hardy HI 1769-FC User Manual

Page 62

Advertising
background image


57

Chapter 7

B1 - Guidelines for Instabilities on Formerly Operating Systems
(Cont’d)

OK?

Yes

No

B1

OK?

Yes

No

Vessel -

When inspecting a vessel, the Center of Gravity (COG)should be
low and centered equally over all the load cells.Insure the load is

directly over or under the load point to avoid side-loading and

that there isn’t any side loading from piping or external forces.

Install flexures on all piping to insure a free floating vessel.
Make sure the vessel and load cell mounts are mechanically stable

and fixed.

Large changes in individual load cells indicate a shift in the COG or
faulty load cells.

Piping and motors can effect the individual load cell readings.
Allow for a higher reading on load cells that support motors and

piping.

Insure pneumatic lines are not applying pressure to the vessel
when energized.

Use check (stay) rods to minimize vessel movement.
Make sure the check rods are loose and not interacting with

the vessel.

Power down all vibration, vacuum and pressurization equipment

during the test process.

Configuration Settings

Incorrect WAVERSAVER settings can cause unstable weight readings.

Adjust to the lowest WS setting that gives you a stable reading.
Higher frequencies with low amplitude vibrations - Use WS settings 1

or 2 lower frequencies with high amplitude vibrations - Use WS setting

3 or higher. Incorrect number of decimal places. Reading weight
increments beyond the equipment applications level. (See Guideline

Calculations below)

Repeatability -

Divide the total load cell capacity, including decimal points by 10,000.
(Expected stable weight reading)

Resolution -

Divide the total load cell capacity, including decimals points by 30,000.

(The amount you can expect to see but not necessarily stable

C

Return to

A

STABILITY

B2

B3

Advertising