#define ERRORADCONVERT 0x0001

#define ERRORADFAILURE 0x0002

#define STATUSINMOTION 0x0040

#define ERRORNOCAL 0x0080

#define ERROREEPROMWRITE 0x0100
// an error occurred when writing to non-
volatile memory

#define NVRDEFAULTED 0x0200 // set
if SETDEFAULTPARAMS command was
given

#define STATUSCHANENABLED
0x8000 // set if channel is enabled

Name

Description

Default

ChanActive

16 bit integer, set to 1 if the
channel is active, 0 if not
active

1

Metric

16 bit integer which
determines the format of
weight values. Metric is the
sum of 3 fields:
1

If bit 7 is set (0x80),
weight is displayed in
Kilograms

2 If bit 6 is set (0x40),

weight is displayed as a
floating point

3. If bit 6 is not set, Weight

is displayed as an
integer, with the 3 least
significant bits giving the
number of decimal
places

0 (weight in pounds, integer
format, 0 decimal places)

WAVERSAVER

16 bit integer
0 No WAVERSAVER
1

4 Hertz

2

2 Hertz

3

1 Hertz

4

0.5 Hertz

5

0.25 Hertz

3 (1 Hertz)

TABLE 4-1: PARAMETERS