Hardy HI 1769-2WS User Manual

Page 55

Advertising
background image

Chapter 4 - Setup

4-21

STABILITYTEST
0x6B
Switch in a specified signal
in place of the normal load
cell signal. With an IT-
JBOX, 4 individual load
cell signals, or a reference
signal on the JBOX may be
switched in.
Without an IT-JBOX, only
an onboard reference signal
may be switched in.
Giving any other command
after STABILITYTEST
causes the unit to return to
normal operation.

O:0 = 0x6B
O:1 = signal to switch in
O = onboard reference signal
1-4 = load cell signals on IT-JBOX
5 = reference signal from IT-JBOX

I:0 = 0x6B
I:1 = COMMAND STATUS

NOTE:
The COMMAND STATUS
INDEXOUTOFRANGE (-4) is
returned if a signal outside the
0-5 range is requested, if you
have an IT-JBOX, or if a signal
other than 0 (onboard refer-
ence signal) is requested and
you do not have an IT-JBOX.

I:2-I:15 See NOCMD,
continuously updating

NOTE:
During the Stability Test,
Gross and Net Weights are
results of signal selected for
this test.

Command

Required Output Table Values
Written by User (PLC)

Input Table Response
From Weigh Scale