Calibration setup procedures, Setting the metric parameter, Setting the motion tolerance value – Hardy HI 1769-2WS User Manual

Page 58: Calibration setup procedures -24

Advertising
background image

HI 1769-WS/HI 1769-2WS WEIGH SCALE MODULE

4-24

TABLE 4-2: COMMAND TABLE

Calibration Setup
Procedures

Setting the Metric
Parameter

The Metric Parameter can be set to either kilograms
or pounds. Any weight value input to the module (e.g.
CALLOWWEIGHT, SPANWEIGHT) are in the cur-
rently selected units. The unit of measure can be set at
any time, not just at calibration. Setting the unit of
measure before calibrating reminds the user what unit
of measure is being displayed. It is important to note
that the weight scale module does not need to be cali-
brated again after changing the unit of measure.

Setting the Motion
Tolerance Value

The motion tolerance is the tolerance value used to
determine if the scale is in motion.

READC2SERIALNUM
0x70
Read data from a C2 sensor.
The C2SEARCH command
must be performed before this
command is done.

O:0 = 0x70
O:1 = SENSOR NUMBER
(0-7)

I:0 = 0x70
I:1 = COMMAND STATUS
I:2-9 = Serial Number
I:10 = Sensitivity, LSW
I:11 = Sensitivity, MSW
I:12 = Capacity, LSW
I:13 = Capacity, MSW

NOTE:
Sensitivity is an integer, with
4 decimal places and dimen-
sions of millivolts per volt.
Capacity is an integer with 0
decimal places, with units of
pounds.

SETDEFAULTPARAMS
0x94

O:0 = 0x94
O:1-15 (unused)

I:0 = 0x94
I:1 = COMMAND STATUS
I:2-I:15 See NOCMD,
continuously updating

EXTENDED VERSION
0xFF
Read full software version
information

O:0 = 0xFF
0:1-15 (unused)

I:0 = 0xFF
I:1 = 0
I:2 = SERIES_REV
I:3 = MAJOR_REV
I:4 = MINOR_REV
I:5 = PATCH_REV
I:6 = BUILD_REV