Zero command (zerocmd), Tare command (tarecmd) – Hardy HI 1756-2WS User Manual

Page 40

Advertising
background image

HI 1756-WS WEIGH SCALE MODULE

4-12

Reply data contains 3 fields

• Command (DINT): The command number
• Channel (DINT): The channel number
• Status (DINT): Status information

NOTE:

Some commands will produce longer replies.

Zero Command
(ZEROCMD)

The Zero Command requests that the current gross
weight be set to zero.

Command Number: 1 (Hexadecimal)
Channel Number: 0 or 1
Number of Elements: 8
Error Return Values:

• OUTOFTOLERANCE - the current weight

value is beyond the limits set by the zero toler-
ance parameter.

NOTE:

The amount of weight zeroed off is cumulative. The
Autozero command will fail if the current gross weight
plus any previously zeroed amount exceeds the zero
tolerance value.

• “STATUSWORD” - the command failed

because either the weight was not stable (in
motion) or there was some kind of A/D error.

Tare Command
(TARECMD)

The Tare Command requests the current net weight be
set to zero.

Command Number: 2 (Hexadecimal)
Channel Number: 0 or 1
Number of Elements: 8

Error Return Values:

• “STATUSWORD” - the command failed

because either the weight was not stable (in
motion) or there was some kind of A/D error.

Advertising