Hard calibration ladder logic example, Hard calibration ladder logic example -7 – Hardy HI 1756-2WS User Manual

Page 59

Advertising
background image

Chapter 5 - Calibration

5-7

For more information on the Cal Low
Command go to Chapter 4, Setup,
page 4-11.

Step 3.

If you used a weight remove it from the
scale.

Step 4.

Place the high (Span) calibration weight
on the scale.

Step 5.

Send a Cal High Command (CAL-
HIGHCMD)
.

The Cal High Command - Sets the
Span Weight parameter to the current
A/D average counts when doing Hard
CAL.

Command Number: 0x65 (Hexadeci-
mal)

Channel Number: 0 or 1

Number of Elements: 8

For more information on the Cal High
Command go to Chapter 4, Setup,
page 4-12.

Step 6.

Remove the weight from the scale.

Step 7.

Send a Write Non-Volatile Command
(WRITENONVOLATILE).

The Write Non-Volatile Command -
causes all parameters (including cali-
bration constants) to be saved to the
non-volatile memory.

Command Number: 4 (Hexadecimal)

Channel Number: 0 or 1

Number of Elements: 8

Hard Calibration
Ladder Logic
Example

Step 1.

Check to be sure that the parameters have
been setup for your weighing process. (See
Chapter 4, Setup)

Step 2.

We have provided a Ladder Logic example
explaining how to set the weigh process
parameters. The Ladder Logic example is