Error code list, Timed out commands, Return codes – Hardy HI 1756-2WS User Manual

Page 49

Advertising
background image

Chapter 4 - Setup

4-21

The least significant 16 bits are a command for Chan-
nel 0, the next 16 bits are a command for Channel 1.

Commands are “1 shot”, occurring upon a 0-1 transi-
tion.

Bit 0x2000 in the STATUSWORD will be set upon
completion of the output table comand. Bit 0x4000
will also be set if the command failed. Setting the
command word to zero will clear these status bits.

Error Code List

SUCCESS

0

OUTOFTOLERANCE

-3

NOSUCHCMD

-5

C2FAILNODEVS

-6

C2FAILCAPEQ

-7

(failure, capacities not
equal)

HARCALFAILCOUNTS -8

(failure, not enough
ADC counts between
high, low)

Timed Out
Commands

The ASIC on the HI 1756-WS module requires that a
command receive a reply within 65 milliseconds.
This is not enough time to complete some of the com-
mands. The commands subject to time-out are:

CALLOWCMD

C2CALCMD

WEIGHSYSTEST

C2SEARCH

The reply to these commands is sent before the com-
mands are actually performed.
To see if a calibration has succeeded, the command
“GETSTATUS” can be sent

The data returned by the WEIGHSYSTEST and
C2SEARCH commands is actually the data deter-
mined by an earlier command. In practice, these com-
mands need to be sent twice.

Return Codes

See Chapter 7, Troubleshooting.