CHAPTER 7 - Troubleshooting

7-3

FIG. 7-2 COMMAND DEFINITIONS AND ACTIONS

Name / Code #

Definition

Action

erroradconvert 1

B1: Load cell input out of range
(i.e., voltage not 0-15 mV and
flashing red LED will display).
Can result from overloaded or
mismounted load cell.In this
state weight readings do not
repond to changes.

C1: Check the voltage levels to
the module from each load cell. +5
V for excitation and sense lines
and 0 - 15 mV on signal lines. If
voltage is bad, to find a problem
load cell, disconnect each one at
the summing box.

erroradfailure 2

B2: Output from the A/D con-
verter to processor is bad. The
module shows a solid red LED.

C2: Contact Customer Support to
return module for repair.

statusinmotion 64

B3: The rate of scale weight
change over 1 second exceeds
the motion tolerance setting. If
the setting is too low, motion
may be indicated when no
changes are occurring.

C3: If the weight is actually
changing, stabilize it. If not,
increase the motion tolerance set-
ting until the motion bit goes off
with static weight.

Status = 1

A1

Yes

C1

Yes

B1

Status = 2

Yes

C2

Yes

B2

Status = 64

Yes

C3

Yes

B3

Status = 256

Yes

C4

Yes

B4

Status = -3

A2

Yes

C5

Yes

B5

Status = -5

Yes

C6

Yes

B6

Status = -6

Yes

C7

Yes

B7

Status = -7

Yes

C8

Yes

B8

Status = -8

Yes

C9

Yes

B9

Definition Action

Definition Action.

.

Advertising
This manual is related to the following products: