Section 3: Setup

11

SECTION 3 SETUP

3.1

SCOPE

Chapter 3 consists of all the procedures to setup the PROFIBUS Interface Option. To
make sure that the interface option works properly, programmers and maintenance
personnel should be familiar with this chapter before setting up or operating the system.

3.2

PANEL, WALL AND REMOTE SETUP PROCEDURE

1. Press the 7/Option button. (See Fig. 3-1) The first option appears.

Figure 3-1

2. Press the “4/Up” ( ) arrow button until ProFi appears on the display. (See Fig. 3-2)

3. Press the “Enter” button. The current node station address appears.

4. Use the keypad to enter a node station address (the valid address range is 1-125). The

station address must be a unique number for each node on the bus.

5. Press the “Enter” button to set the node station address.

NOTE:

The PROFIBUS node address number is displayed in decimal on the weight controller.