Section 6: Selectable Reads

43

6.7.10 INSTRUMENT ID

6.7.11 INSTRUMENT IDENTIFICATION

6.7.12 RESPONSE/ERROR CODE

Instrument ID

# Words

#

Bytes

Parameter

#

Hex

Number

Tare Weight
Net Weight
Gross Weight
Print Initiation (0 = Continuous, 1 = Altered
print)
Reserved for future use (must be set to 0)

Baud Rate (0 = 600, 1 = 1200, 2 = 2400,
3 = 4800, 4 = 9600, 5 = 19200)

Port Configuration (must be set to 1for printer

output only)

Stop Bits (0 = one stop bit, 1 = two stop bits)
Parity (0 = none, 1 = even, 2 = odd)
Handshake Control (0 = hardware, 1 = software)
Word Length (0 = seven bits, 1 = eight bits)
Device Address (a value from 0-99)
Echo (0 = off, 1 = on)

bit 3
bit 4
bit 5
bit 6

bit 7







0.5

0.5

0.5
0.5
0.5
0.5
0.5
0.5






1

1

1
1
1
1
1
1






91

92

93
94
95
96
97
98






5B

5C

5D
5E

5F

60
61
62

Instrument Identification

#

Words

#

Bytes

Parameter

#

Hex

Number

Firmware Revision Level (one character text, i.e. A)

(Pursuant to Revision Specification ASME
Y14.35M-1992)

Firmware Version (ASCII text number)
Zero Calibration analog-to-digital converter raw
counts
Span Calibration analog to digital converter raw
counts

0.5


2
2

2

1


4
4

4

100


101
102

103

64


65
66

67

Response/Error Code

#

Words

#

Bytes

Parameter

#

Hex

Number

Provide the last write command number with its
response/error code. (See Chapter 8 for a list of
response/error codes)

1

2

0

0