KEYENCE EX-V Series User Manual

Page 36

Advertising
background image

Chapter 2 Quick Guide to Basic Mode Operation

2

26

3. Press the

key.

The sensor automatically enters the HIGH limit value setting mode.

The setting value is displayed in red and the HIGH indicator illuminates.

Note: Setting the panel-lock function to ON allows the terminals to switch the
setting number for the tolerance limits. For details of the panel-lock function, see
"Panel-lock function [ ]" (

➮ page 43).

4. Press the

or

key to display the desired value.

Pressing the

key changes the flashing digit to the right.

Pressing the

key changes the value.

All digits flash when the fifth digit and sign can be changed.

5. Press the

key.

The HIGH limit setting value is saved. The sensor automatically enters the
LOW limit value setting mode.

Set the LOW limit value using the same procedure as in steps 3 and 4.

Note 1: The output is disabled during the tolerance limit value setting. When the
output is already on, however, the output is retained.

Note 2: Pressing the

key does not save the changes if [HIGH setting value -

Hysteresis] is smaller than [LOW setting value + Hysteresis]. In this case, “Err1” is
displayed. Press the

key again to cancel the error, and then specify the correct

values.

HIGH

TIM

GO

LOW

HIGH

TIM

GO

LOW

HIGH

TIM

GO

LOW

No setting

(5th digit)

(1st to 4th digits)

W