Chapter 3 Use of Common Functions

3

37

5. Set the display value for correction point 1.

The display value is displayed in red.

• The value on the flashing digit can be changed.
• Pressing the

key changes the flashing digit to the right.

• Pressing the

key changes the value.

• All digits flash when the fifth digit and sign can be changed.

6. Press the

key to save the display value for correction point 1.

• Then, the sensor enters the monitor output value setting state.

7. Set the monitor output value for correction point 1 in the same procedure as

step 5, and use the

key to save the setting value.

• The monitor output value is displayed in green.

8. Repeat steps 5 through 7 to set and save the display and monitor output values

for correction point 2.

After saving the monitor output value for point 2, the sensor returns to the
measurement state.

Note 1: The display and monitor output values for monitor output scaling can be
set between -19999 and +19999.

Note 2: To set the scaling function, be sure to satisfy the following conditions. If
these conditions are not satisfied, "Err3" is displayed and the setting procedure is
disabled. When the error message is displayed, press the

key to cancel the

error message and retry the setting.
(1) Display value 1 - display value 2

≠ 0

(2) Monitor output value 1 - monitor output value 2

≠ 0

(3)

No setting

(1st to 4th digits)

(5th digit)

0.001

(monitor output value 2 - monitor output value 1 )

≤ 6

(display value 2 - display value 1)