KEYENCE EX-V Series User Manual

Page 45

Advertising
background image

Chapter 3 Use of Common Functions

3

35

No setting

(1st to 4th digits)

(5th digit)

When "In-1" is selected

5. Set and save the input value for correction point 1 using the value measured by

the sensor (input from the sensor head).

Move the sensor head or target to the distance to be set as correction point
1.

Press the

key to save the measured value.

6. Use the

or

key to set the display value for correction point 1.

Pressing the

key changes the flashing digit to the right.

Pressing the

key changes the value.

All digits flash when the fifth digit and sign can be changed.

7. Press the

key to save the setting value.

8. Set and save the values for correction point 2.

Repeat steps 5 through 7 to set and save the input and display values for
correction point 2.

After saving the display value for point 2, the sensor returns to the
measurement state.

Note 1: The input and display values for display scaling can be set between -
19999 and +19999.

Note 2: To set the scaling function, be sure to satisfy the following conditions. If
these conditions are not satisfied, "Err2" is displayed and the setting procedure is
disabled. When the error message is displayed, press the

key to cancel the

error message and retry the setting.
(1) Input value 1 - input value 2

≠ 0

(2) Display value 1 - display value 2

≠ 0

(3) After scaling, the display value is within the range of

±19.999 when the input

value is "0".

(4)

Note 3: When using the internal trigger function (

➮ page 46), adjust the trigger

level after performing scaling. (The internal trigger function works with regard to the
value displayed after scaling is performed.)

0.0001

(display value 1 - display value 2 )

≤ 15

(input value 1 - input value 2)

Advertising