quIcK conFIguraTIon procedure

1. Apply power to the unit while pressing the enter key. The display shows:

QUICK COnFIG

PRESS EnTER...

...TO COnTInUE

2. Press the enter key. The display shows:

UnITS? lbs

The choices for units are lbs or kg. For kg, press the up Arrow key; the display shows:

UnITS? kg

note: you can change units after Quick Config is complete.

3. When the display shows the desired units, press the Enter Key. The display acknowledges the selection

and shows:

km BOLT-On?

The choices for sensor type are K-M bolt-on or direct support. For direct support sensors, press the Up
Arrow key;the display shows:

DIRECT SUPPORT?

4. When the display shows the desired sensor type, press the Enter Key. The display acknowledges the selection.

• For km bolt-on sensors, proceed to Step 5.

• For direct support sensors, proceed to Step 16.

5. If you selected km Bolt-On, the display shows:

L-CeLLs?

The choices for km bolt-on sensor are L-Cells or microcells. For microcells, press the up Arrow key; the
display shows:

mICrOCeLLs?

6. When the display shows the desired bolt-on sensor, press the Enter Key. The display acknowledges the

selection.

• For L-Cells, proceed to Step 11.

• For microcells, proceed to Step 7.

7. If you selected microcells, the display shows:

HOrIZOnTL BeAmS?

The choices for location are horizontal beams or vertical legs. For vertical legs, press the up Arrow key;
the display shows:

verTICAL LegS?

8. When the display shows the desired Microcell location, press the Enter Key. The display acknowledges

the selection.

• For microcells on horizontal beams, proceed to Step 11.

• For microcells on vertical legs, proceed to Step 9.